Switch |
Name and Switch Setting |
01 |
Beeper Tone |
|
00 - off |
|
01 - soft |
|
02 - medium |
|
03 - loud* |
04 |
RS232 Handshaking |
|
00 - off* |
|
01 - cts |
|
02 - busy |
|
03 - xon |
05 |
Auto transmit |
|
00 - off* |
|
01 - on |
06 |
Decimal point |
|
00 - off |
|
01 - on* |
07 |
Response data separator modes |
|
00- space |
|
01 - comma* |
|
02 - tab |
|
03 - carriage return |
|
04 - carriage return & line feed |
|
05 - line feed |
08 |
Response delimiter in RCI mode (see 12CF) |
|
00 - carriage return |
|
01 - carriage return & line feed* |
|
02 - line feed |
09 |
Echo settings |
|
00 - off* |
|
01 - on |
0A |
Decimal Precision |
|
00 - . |
|
01 - .0 |
|
02 - .00* |
|
03 - .000 |
|
04 - .0000 |
|
05 - .00000 |
0F |
Show data labels |
|
00 - off* |
|
01 - on* |
12 |
RS-232 Protocol modes |
|
00 - RCI* |
|
01 - ICP |
13 |
Measurement modes |
|
00 - Static measurement mode |
|
01 - Dynamic measurement mode* |
14 |
Number of samples to average per static reading |
|
0x01* |
|
. |
|
. |
|
0xff |
15 |
Number of readings taken in static measurement |
|
0x01* |
|
. |
|
. |
|
0xff |
16 |
Illuminant/Observer types |
|
00 - A_2 |
|
01 - A_10 |
|
02 - C_2 |
|
03 - C10 |
|
04 - D50_2* |
|
05 - D50_10 |
|
06 - D65_2 |
|
07 - D65_10 |
|
08 - F2_2 |
|
09 - F2_10 |
|
0A - F7_2 |
|
0B - F7_10 |
|
0C - F11_2 |
|
0D - F11_10 |
|
0E - F12_2 |
|
0F - F12_10 |
|
------------- |
|
10 - D75_2 (added after firmware v8212) |
|
11 - D75_10 (added after firmware v8212) |
|
12 - D55_2 (added after firmware v8212) |
|
13- D55_10 (added after firmware v8212) |
17 |
Density status types |
|
00 - Status_A |
|
01 - Status_E |
|
02 - Status_I |
|
03 - Status_T* |
|
04 - Spectral |
|
------------ |
|
05 - Spectral_X (added after firmware v8212) |
|
06 - Status_HIFI (VCMYRGBO) (added after firmware v8212) |
|
08 - Status_Tx (added after firmware v8212) |
|
09 - Status_Ax (reflectance) (added after firmware v8212) |
|
0A - Status_Ex (added after firmware v8212) |
|
0B - Status_G (added after firmware v8212) |
|
0C - Status_M (added after firmware v8212) |
|
0D - Status_Axt (transmission) (added after firmware v8212) |
|
0E - Status_Mx (added after firmware v8212) |
Note: Status responses with the sub-x option are responses with correction factors entered to more closely match the existing population of filtered densitometers. Status responses without the sub-x option correspond directly to the ANSI definitions. |
18 |
Data Output types |
|
00 - Density |
|
01 - Plus Dot |
|
02 - Minus Dot |
|
03 - Reflectance* |
|
04 - Lab |
|
05 - XYZ |
|
06 - Yxy |
|
07 - xyY |
|
08 - LCh |
|
09 - Density (VCMYRGBO) |
19 |
Instrument mode (added after firmware v8212) |
|
00 - Reflectance* |
|
01 - Transmission |
|
|
|
Additional Commands |
CM |
Calibrate Drive motor |
CR |
Calibrate Reflectance |
|
This executes reflection calibration using the supplied calibration reference |
CU |
Calibrate Unit (added after firmware v8212) |
|
This enters the CR, CM and, if transmission mode is present CT |
DM |
Drive Motor - turns on/off the drive motor |
|
02 DM returns current motor speed |
|
1.20DM will set the motor to the default motor speed |
RI |
Reset Instrument to its power - up state |
RM |
Request Measurement Initiates the measurement process |
SV |
Send Version returns the model and firmware version |
XE |
Clear and transmit errors |